Hi all,

It might be a good idea to confirm part numbers as well as ECL levels. I think
at some point they added additional functionality to the 2H-TX and changed the
part number. If you had a newer one and swapped in an older one it could cause
loss of some functions. You might be able to find the part number in the BSMS
tool, or just pull the board out and check the part number on the sticker.

Cheers,
Mike

> -----Original Message-----
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Karel Klika
> Sent: Tuesday, May 30, 2023 11:26 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] AMMRL: lock problem
>
> P.S.
>
> Our board is BSMS/2, ECL01. Not sure what version it could go up to, e.g. we
> have SGU's ECL01, ECL06, and ECL07.
>
> Many thanks to Martha for the prompt.
>
> Regards,
> Karel
>
> ________________________________________
> From: Martha Morton <mmorton4_at_unl.edu>
> Sent: 30 May 2023 19:58
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] AMMRL: lock problem
>
> Karel,
>
> Please tell us your ECL level for the 2H-TX board.
>
> Martha
>
> Martha Morton
> Director of Research Instrumentation
> Co-Chair, Department of Chemistry Safety Committee
> University of Nebraska-Lincoln
> 4024726255<tel:4024726255>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#159): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/159__;!!PvDODwlR4mBZyAb0!TdznOciYqkZakUpCr6-ANz_QZpbfKE8n=
HUsJ5iZPD7uxde1IsGoSCQr9kJnkTyOGHjAuQ8KuMNLZoRP0TA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99149994/=
7559972__;!!PvDODwlR4mBZyAb0!TdznOciYqkZakUpCr6-ANz_QZpbfKE8nHUsJ5iZPD7uxde=
1IsGoSCQr9kJnkTyOGHjAuQ8KuMNLdGLrK9Q$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Tue May 30 2023 - 11:20:58 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:41:33 MST