Hello Karel,

What kind or probe are you using? Is it a cryoprobe or RT probe?
The figure you sent of your attempt to measure the deuterium pulse length
looks like you are just seeing artifacts of some kind. I would say that
you might be having a problem with the preamp slice. But you can lock,
so I'm not sure what is happening here. Maybe something to do with the
transmit/receive switch in the preamp?

Have you had a look at the wobble curve on 2H?

Bye for now,
Bob.

-----Original Message-----
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Karel Klika via groups.io
> Sent: May 26, 2023 6:26 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] AMMRL: lock problem
>
> Dear all,
>
> We have a problem with lock/topshim. We can lock fine on 2D, lock is normal
> and S/N is good. If we run topshim on 1H all is fine. However, topshim on
> 2D fails and we get an error message saying no signal at all. Measuring a 2D
> spectrum only provides a very weak signal. If I try and measure the 2D pulse
> length it looks really weird (pic attached), the signal intensity does not
> seem to change at all, but the phase is totally erratic. Any ideas what
> the problem is and how to fix it ?
>
> Many thanks.
>
> Regards,
> Karel

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#149): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/149__;!!PvDODwlR4mBZyAb0!T4tqgfZbSc1sSQtRvR0MAZkSN923cC5F=
wBN5LYE-y1iCOG-Ag3L4EAV6NFefdqCc3z6FcdliJ0lAEXpZ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99149994/=
7559972__;!!PvDODwlR4mBZyAb0!T4tqgfZbSc1sSQtRvR0MAZkSN923cC5FwBN5LYE-y1iCOG=
-Ag3L4EAV6NFefdqCc3z6FcdliJ4ZTE3Er$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Fri May 26 2023 - 07:43:11 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:41:33 MST