Ken,
The others have already listed several good things to note, so I'll add a
couple of suggestions.
If you absolutely need chilled water, it's probably a good idea to get some
Haskris or Neslab portable water chillers that can handle the thermal output
of your equipment, since chilled water lines will probably be shut off on
a regular basis.
For HVAC, the portable air conditioners can help, but again, you'll need
multiple units for that.
You'll have to specifically point out what plumbing must remain untouched,
especially for your air compressor and the piping. It's not entirely unusual
for the wrong pipes to somehow get shut off, etc., if specific instructions
aren't given.
If your institution insists on moving your facility, then remind them that
the sheer cost of liquid helium (de-energizing the magnets, warming them up,
transporting them, and cooling them down again) will be prohibitive.
Ronald Shin
University of Alabama at Birmingham
Central Alabama High Field NMR Facility
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Kenneth Sharp-Knott via groups.io
> Sent: Wednesday, May 17, 2023 9:23 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Upcoming Building Renovations - Planning
> Have any of you undergone major building renovations and how did you handle
> this in your NMR facility? I'm would like some advice from those who have
> gone through this process.
> I'm unclear just how gutted the building is planned to be, but I believe it
> will be a fairly major renovation including HVAC and chilled water, etc.
> so I'm not sure how the facility can coexist during the process.
> This is many years ahead, but I would like to advise the department on the
> impact to the NMR facility which obviously can't just be moved to another
> temporary location.