AMMRL: What is QNP Temperature range? -ANSWER +180 to -150 C

From: Hirschinger, Jerry D <jerryh56_at_purdue.edu>
Date: Wed, 18 May 2016 22:16:28 +0000

Many thanks to all who responded (Abil, Clemens, George, Bob, Jane, Joe, Aleksej, Linda, Charlie, Erika).

The VT range for the probe is -150 to +180 C. One must be careful to use Blue plastic spinners only from RT to 80C, White plastic spinners up to 100C, and ceramic spinners outside of these ranges. Ceramic works better at low temps because the added weight helps keep the sample down under higher VT airflow.
Cheers!!


Hello Spinland!

We are planning some variable temperature work on our old ARX400 with a QNP, but I cannot find the published allowable temperature range of the QNP. This is a 1990 vintage 400MHz QNP. Does anyone have any documentation of the allowable temperature range for this QNP?

Thank you, -Hirsch