AMMRL: Temperature calibration

From: Keith Brown <brownk_at_chem4823.usask.ca>
Date: Thu, 20 May 2010 13:04:54 -0600 (CST)

   Hi all. I have a student who is doing a lot of low and high temperature
work. Part of her study involves looking at the accuracy of the
temperature readings that the probe thermocouple is providing. She is
using 4% methanol in methanol-d4 with a trace of HCl for low temperature
calibration and 80% 1,2-ethandiol in DMSO-d6 for high temperatures (from
Berger's 150 and More NMR Experiments).

   Not surprisingly, perhaps, she is finding that as the temperature is
decreased from room temperature the actual temperature is lower than the
reading given by the thermocouple. Furthermore, delta T (where delta T =
Treading - Tcalc) increases as the temperature is decreased. A plot of
delta T vs Treading is linear (mostly) for low temperatures. For high
temperatures, using 1,2-ethandiol, the situation is reversed. delta T in
this case is always negative, as one might expect given the linear nature
of the low temperature plot (which reaches zero at 298K). However, there
are two problems with the high temperature calibration plot. First, it
isn't linear from 298K to 370K but rather, curves down a bit and then up
to zero. Secondly, at 298K there is a large discontinuity in the
temperature calibration plot. Methanol gives a delta T of about +1 degree
and 1,2-ethandiol gives a delta T of -5 degrees.

   So, finally my questions. Why the non-linearity and why the large
discontinuity? What are your experiences and techniques used for
temperature calibration?

   Cheers :)

Dr. Keith Brown
Department of Chemistry/
Saskatchewan Structural Sciences Center
University of Saskatchewan
Saskatoon, Saskatchewan
306-966-1725
http://chem4823.usask.ca/kbrown.html

-----------------------------------------------------------------------
If you're not part of the solution, you're part of the precipitate.
-----------------------------------------------------------------------
Received on Thu May 20 2010 - 09:05:17 MST

This archive was generated by hypermail 2.4.0 : Fri Jun 16 2023 - 16:55:17 MST