Hello!
When you do the charging of magnet and cryoshim it, it will see the
surrounding magnetic paths and accordingly will shim out. I mean, the
cryoshimming will nullify the effect of magnetic(Non-moving) material.
Regards.
L S Mombasawala.
NMR ab,
SAIF, IIT Bombay.
_____________________________________________________________
Hi all,
We are planning to move an old 600MHz non-shield magnet to a basement but
worry about the crossing metal pipe lines on the ceiling which will fall
into the 5 gauss line. These pipe lines are definitely NOT moving and we
assume they will not affect the homogeniety of the magnet after shimming
adjustment. Does anybody have an experience to deal with this situation?
Any input will be highly appreciated!
Best regards,
Guangyu
Received on Fri May 04 2007 - 17:44:40 MST