Pt 195 NMR

From: Wei Li <wli_at_utmem.edu>
Date: Sat, 07 May 2005 10:27:30 -0500

Dear friends,

A few weeks ago I asked for suggestions on running
Pt-195 NMR using Varian's switchable probe, while I
was waiting for the standard to arrive. I got a lot
of responses and all of them are very informative.
With K2PtCl6 in water and a trace of HCl as
standard, use a little bit of back linear
prediction, the nice singlet peak appears with very
flat baseline. Thank you all for all the nice
suggestions and if anyone need a workable parameter
file for Varian-500, please let me know.

Regards,

Wei.

------------------------------------------
Wei Li, Ph.D
Assistant Professor
College of Pharmacy
University of Tennessee Health Science Center
847 Monroe Ave., Room 327A
Memphis, TN 38163
(901)4487532 (Tel)
(901)4486828 (Fax)
Received on Mon May 09 2005 - 12:38:24 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 10 2023 - 17:44:25 MST