RE: Warning notice

From: Banks, Benjamin N. <benjamin.banks_at_targacept.com>
Date: Fri, 28 May 2004 08:01:39 -0400

That's essentially the way we would do it at former NMR sites. When the
room needed cleaning or maintenance of some kind, we just arranged to have
someone there to oversee them after carefully explaining things to them.
Otherwise, the room was off limits.


Benjamin N. Banks, Ph.D.
Targacept, Inc.
200 East First St, Suite 300
Winston-Salem, NC 27101-4165
(336) 480-2234 Phone
(336) 480-2113 FAX
benjamin.banks_at_targacept.com
www.targacept.com

CONFIDENTIALITY NOTE: This e-mail message, including any attachment (s),
contains information that may be confidential, protected by the
attorney-client or other legal privileges, and/or proprietary non-public
information. If you are not an intended recipient of this message or an
authorized assistant to an intended recipient, please notify the sender by
replying to this message and then delete it from your system. Use,
dissemination, distribution, or reproduction of this message and/or any of
its attachments (if any) by unintended recipients is not authorized and may
be unlawful.



-----Original Message-----
> From: Neil Jacobsen [mailto:neil_at_u.arizona.edu]
> Sent: Thursday, May 27, 2004 4:03 PM
> To: ammrl_at_ammrl.org
> Subject: Re: Warning notice


I think it would be worth saying something about cleaning people. We don't
allow any cleaning people in any of our NMR rooms. Of course that means we
have to empty the garbage ourselves, and the rooms are filthy, but I think
it's worth it. I think this policy started when I was trying to explain
the magnet situation to a new custodial staffer who was deaf, and after
awhile it dawned on me that maybe we didn't really need to have the rooms
cleaned. Of course, this policy has to be explained to the supervisors
since staff changes regularly and all of these people have master
keys. We've also had problems with maintenance people coming into the
rooms to turn valves, shut off power, etc., and you just have to be
vigilant. I try to get to know the maintenance people and have them always
go into these rooms with a "chaperone".

Neil


At 05:13 AM 5/27/2004, you wrote:
>Hi,
>I had someone get a metal chair stuck to the base of a 500 mhz magnet last
>year. It took 3 people and a few 2x4's to break it away slowly. No quench.
>Lucky. Field came back and is stable.
>We did have some notices posted on the doors and the magnet. However the
>cleaning person couldnt speak English nor able to read much of it let
>alone technical terms.
>Since then I have warnings posted in several languages. ie; JAK
MAGNET!,etc/
>
>JIm
>
>Alan Boyd wrote:
>
>>Dear AMMRLers,
>>
>>We had a safety inspection this morning, and it was pointed out that in our
>>rather small open-access lab there was no warning notice about magnet
>>quenches. I said that this was a low probability occurrence, and it usually
>>happened when someone knowledgeable was fiddling about with the magnet
>>anyway.
>>
>>We were just about to agree to put up a plain text warning to say 'get out
>>fast if there's a quench', when someone remarked that it was quite possible
>>for an enthusiastic cleaner's floor-polishing machine to get sucked into
>>the magnet, and then the probability of a quench was quite high... as is,
>>unfortunately, the probability of the cleaner being unable to read.
>>
>>So, has anyone devised or designed a graphical or cartoon warning sign that
>>tries to say 'get out fast if there's a quench'?
>>
>>Just trying to avoid re-inventing the wheel...
>>
>>Alan
>>

Neil E. Jacobsen, Ph.D.
NMR Facility Manager
Department of Chemistry
119 Old Chemistry
1306 E. University
University of Arizona
Tucson, AZ 85721
520-621-8146
FAX 520-621-8407 setup from
Received on Fri May 28 2004 - 11:04:41 MST