Sandip Sur (sur@chem.rochester.edu)
Tue, 16 Mar 1999 10:38:33 -0400

Hi Chuck,

I have been using the Kermit on the QE/1280 and the versaterm software on
the powerPC using a serial line with only 3 pins. I had to lower the baud
rate on the powerPC to 9600. Try lowering the baud rate first. Since you
had a working file transfer protocol, the change you need now for powerPC
must be very simple.

Sandip

>We have been using the Tecmag FILTRN program to move files off the 1280
>computers in our QE's. I have recently upgraded one of our mac's from a
>Centris to a PowerPC 7100. Now when I try to transfer the files the
>PowerPC locks up. The line is good because I can send a message from the
>PowerPC to the 1280 with no problem. I really don't want to give up this
>PowerPC but if I can't move file, it is of no use to me.
>
>The Mac had the following message "message: 1st secs recieved"
>The QE gave the following Error #12 BKGRD File transfer local error-use TS
>command.
>
>The TS command say it timed out transfer aborted.
>
>Is it possible there is a timing problem between the 1280 and the PowerPC?
>The Baud rate is set to 19200.
>
>
>Chuck
>
>
>Charles L. Anderson, Ph.D.
>Manager NMR Facilities
>Department of Chemistry
>University of Houston
>Houston, TX 77204-5641
>Office: (713)-743-2728
>Fax: (713)-743-2709
>
>"Anything one man can imagine
> others can make true"
> Jules Verne

________________________________________________________________________________
Sandip K. Sur, Ph.D. Email: sur@chem.chem.rochester.edu (W)
NMR Facility
Department of Chemistry Tel.: (716) 275 4705 (W)
B53B Hutchison Hall (716) 275 0912 (W)
River Campus (716) 586 7794 (H)
University of Rochester Pager: (716) 275-4231 (ask for me)
Rochester, NY 14627-0216 Fax: (716) 473 6889

Tuesday, March 16, 1999
________________________________________________________________________________