Re: AMMRL: Iconnmr Error on opening an existing *.set file, TopSpin 4.1.1

From: Abil Aliev <a.e.aliev_at_ucl.ac.uk>
Date: Tue, 23 Mar 2021 10:27:23 +0000

"Ti" was included accidentally in the list of user-specific parameters
which led to this error!

Clemens Anklin pointed to this.

Thank you,

Abil

On 20/03/2021 23:42, Abil Aliev wrote:
> The set file of Feb16 which failed to open had some title lines which
> did not make sense, e.g. the title line for user ea with the sample in
> position 46 appeared for the sample submitted by user vmc in position 19:
>
> ##HolderIdent## 19 {} {} 19
> ##Holder## 19 Submitted NA-CO-PAPER 12 {CDCl3        chloroform-d} {n
> C13CPD.ucl          13C spectrum} {D:\500} 03:35:06 {UnMoonDisabled
> {ParCLockNPrintDisabled {d8 0 o1 15090.9 o2 2000.52 ns 4096 rg 101 ro
> 20 ti {
> C13CPD.ucl CDCl3 {D:\500} ea 46
>
> } sr -15.76 isen 30 o1p 120.000 o2p 4.000 bcfw 1 aunm au_zg locnuc
> 2H}} {} {} {} {}} vmc {19.2308 3 1} {} {{} {} {} {} AN_OFF {}} 0 {} {}
> {{} {{} {} {} {} {} {} {} {} {} {} {} {}}} {}
> 707e1177-856c-41dd-9895-589cc2cd72e3
>
> The same title "C13CPD.ucl CDCl3 {D:\500} ea 46" appeared in the new
> set file of Mar19 for position 1 (!):
>
> ##HolderIdent## 1 {} {} 1
> ##Holder## 1 Running BT100 20 {MeOD         methanol-d4} {n
> C13CPD.ucl          13C spectrum} {D:\500} 03:35:06 {UnMoonDisabled
> {ParCLockNPrintDisabled {isen 30 o1p 120.000 o2p 4.000 locnuc 2H d8 0
> o1 15090.9 o2 2000.52 ns 4096 rg 101 ro 20 ti {
> C13CPD.ucl CDCl3 {D:\500} ea 46
>
> } sr -15.76 aunm au_zg bcfw 1}} {} {} {} {}} hch {1.193 1 1} {} {{} {}
> {} {} AN_OFF {}} 0 {} {} {{} {{} {} {} {} {} {} {} {} {} {} {} {}}} {}
> a2668b0f-729b-4c11-afdc-5375858ab29e
>
>
> Having title lines with an empty line (between ea 46 and } above) did
> not look right to me. I replaced these with "ti {}" in Feb16*.set and
> this set file can now be opened successfully.
>
> I then checked <date>.set files under TS 4.0.6 and 3.6.2 on other
> spectrometers where we do not have any issues with opening old set
> files. In those set files there was no occurence of "ti " (ti followed
> by space) at all.
>
> Thank you,
>
> Abil
>
>
> On 19/03/2021 22:16, Abil Aliev wrote:
>> Hello,
>>
>> On opening the existing <date>.set file with the list of queued
>> experiments after restarting Topspin and Iconnmr, an error is issued:
>> "Error:unmatched open brace in list..."
>>
>> Has anyone come across a similar error? Is there a fix for this?
>>
>> Thank you,
>>
>> Abil
>>
>>
Received on Tue Mar 23 2021 - 00:27:10 MST