Dear Lab Managers,

At ENC one of the discussions was about system monitoring, routers and power outages. Several of you described situations where backup systems failed during power outages. The system was supposed to send a signal if it loses power, which it does because it has a UPS. The signal isn't received because the router is not on back up power as well. This is the bane of our existence. We don't always control how signals are routed. One of the Bruker people suggested instead of monitoring for a distress signal, the system should be set up so there is always a ping. The loss of the ping is the signal. I hope this helps several of us as we design back up systems for the lab.