Hi Alessia
Though my sustained expertise is solutions NMR, I wanted to add a few
pointers that might be useful in your arsenal of troubleshooting.
I presume you are spinning the samples at the same rate in the above two
spectra, just as a sanity check.
I hope you have ruled out the processing related cause of baseline roll
such as a large or n*360 value for 1st order phase value. Kindly bear with
me and I am once again assuming that you are processing these data in the
phase sensitive mode i.e. not a magnitude or a power spectrum.
If the above is true, then my next suspicion falls on the deadtime because
that is the one thing that can have immediate effect on the baseline
quality. So, have you tried to increase the DE and see ? If the baseline
distortion is caused by the ring down being larger (this indeed can be
related to the preamplifier), even if I process the FID in power spectrum
or magnitude mode, I will expect the distortion to be there. Another test
could be this. If you can record a 'solid echo', then you can digitize the
data from the true peak position of the echo. Such a raw data should not in
principle need any phase correction at all for both the zero and first
order. A FT of such raw data should give a flat baseline, without even any
interactive phase correction. Assuming that the FID is long enough, the
above echo experiment should completely remove the dead time ringing from
the picture. If you still see the distortion, then you can move to the
next step of your troubleshooting i.e. you can preclude the dead time
ringing as the source of your problems.
So, what next step one can think of, if ring down is not the issue here ?
I noticed with interest, you mentioning the increase in pulse width by a
factor of 2.5 approx. A factor of 2 increase in pulse width normally
translates to a factor of four reduction in the total RF power getting in
to the probe. So, the best thing to do is to actually measure with an
oscilloscope and suitable attenuator load (30 dB or above), how much power
is going into your probe. You might do this both with and without the
preamplifier along the path. Normally, older preamps of Bruker i.e. HPPR
vintage quote an insertion loss of about 1 dB. The younger generation
HPPR/2 slices boast an insertion loss as low as 0.1 dB ! So, if you see
much larger insertion loss, you can possibly pin down the problem to the
preamp.
Last, you said you also switched the power amplifier along with the
preamplifier. You might try to change one variable at a time, if feasible,
by keeping the Amp same and try the different preamps and check the above.
Good luck with your troubleshooting !
Best Regards
Rajan
On Wed, Mar 25, 2020 at 1:22 PM Lasorsa, A. <a.lasorsa_at_rug.nl> wrote:
> Dear all,
>
> Hope everybody is doing well in these very bad times and that things will
> get better soon.
> In the meantime, I hope you can help with the following.
> Recently we have faced a new (baseline) problem in recording 1D Sn NMR
> spectra, working on a H-X MAS probe.
> On the same probe, previously, we have recorded analogous spectra (i.e.
> zg30, hpdec with 1H decoupling) with quite satisfying results.
> However, after one technical intervention, where the preamp configuration
> was changed (we switched from a 1000W to a 500W preamp on X, and we also
> changed routing, ie we exchanged X-channel from TRX3 to TRX2), we are
> encountaring some troubles. In particular, the baseline looks like very
> distorted. Also, we are reporting a lower power efficiency, as the 90
> degree pulse seems to be increased from 3.8us to 9us (at the same power,
> 150W).
> The acquisition parameters are exactly the same as before, large spectral
> width, long dead time and short dwell time. I have attached you two
> screenshots showing spectra before and after. Of course this is the best
> scenario, using a "standard" sample (SnI2) with a good si/no ratio. Moving
> to different samples, things get worse and signals are hidden in this sort
> of ringing pattern.
>
> Thanks in advance.
>
> Best wishes
> Alessia
>
--
____________________________________
Rajan K Paranji, Ph.D.NMR Facility Manager
*Department of Chemistry**Room 65, Bagley Hall*
University of WashingtonSeattle, WA 98195
phone : 206 685 2581 fax: 206 685 8665email: rajanp_at_uw.edu
____________________________________
*PLEASE NOTE: I will be available in person on Monday, Wednesday and
Friday forenoon*
*to attend to your needs. On other days I will be 'teleworking' and
you can reach me *
*via email...Thank you for your understanding*
Received on Wed Mar 25 2020 - 11:04:07 MST