From: Carlos Amezcua <Carlos.Amezcua_at_UTSouthwestern.edu>
Date: Fri, 23 Mar 2007 15:18:42 -0500

Hi all,
        Does anyone out there knows an easy way to measure the Molybdate/
Reactive-Silicate concentration in water?. The new Sumitomo helium
compressor that Varian is beginning to ship with the new cryogenic
probes requires a concentration of < 50mg/L of this material in the
cooling water. Also, can anyone recommend a brand of test strips for
measuring water hardness in mg of CaCO3/L?

Thanks in advance for your help