Re: [AMMRL] π/2 observe pulse for Bruker HRMAS probe #Bruker

From: Apostolos Spyros <aspyros_at_uoc.gr>
Date: Wed, 28 Jun 2023 12:17:48 +0300

Our 1.5 year old 4 mm HRMAS iProbe (400 MHz) has the following:

1H: 5.0 us _at_ 10.8 W

13C: 6.0 us _at_ 73.0 W

A.

On 27/6/2023 8:40 μ.μ., Fu Chen wrote:
> Hi all,
> Can anyone share the parameters of π/2 observe pulse length and power
> level (1H, 31P, 13C and 15N) for Bruker 4mm HRMAS probe? The better
> the most recent probe.
> Thanks very much!
> Fu
> --
> Fu Chen, PhD
> Director of Analytical NMR Facility
> Department of Chemistry and Biochemistry
> B0208A Chemistry Building
> University of Maryland, College Park, MD 20742
> Tel: 301-405-1869
> fchen127_at_umd.edu
> https://blog.umd.edu/nmr/
>

--
A. Spyros
Assoc. Professor in Analytical Chemistry
_____________________________________________________

NMR Laboratory
Department of Chemistry
University of Crete
Voutes campus
Heraklion Crete 70013
Greece

office: +30 2810 545085
lab: +30 2810 545112
fax: +30 2810 545001
email:aspyros_at_uoc.gr
web:http://www.chemistry.uoc.gr/aspyros
skype: apostolos.spyros
_____________________________________________________


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#248): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/248__;!!PvDODwlR4mBZyAb0!W3EW6Ih_d3DgeqqpOLs5nRH0E0UDS87Q=
dpBUGzRAxjosST2L_rFpj5X-BL0r2uKrqrggzRhG3pJnLEM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99814285/=
7559972__;!!PvDODwlR4mBZyAb0!W3EW6Ih_d3DgeqqpOLs5nRH0E0UDS87QdpBUGzRAxjosST=
2L_rFpj5X-BL0r2uKrqrggzRhGtVPc0dQ$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!W3EW6Ih_d3DgeqqpOLs5nRH0E0UDS87QdpBUGz=
RAxjosST2L_rFpj5X-BL0r2uKrqrggzRhGsEZhM4w$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jun 28 2023 - 02:18:08 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST