Re: [AMMRL] NMR relocation

From: krishnarao doddapuneni <krishnaraod_at_tifrh.res.in>
Date: Tue, 27 Jun 2023 11:56:00 +0530

Hi Tianzhi,

We have relocated our two 700 and two 500s and 300 Bruker spectrometers. I
would recommend using a truck with very good suspension and fully covered
to avoid road vibrations and protect from rains. I would prefer to do
proper deenerging rather than just just quench the magnet and re energise.
Also recommend to keep small screws/spares packed in a well recognisable
pockets/bags.

I wish you all the best. We are not sure the reason but our Ascend 500 gave
trouble after relocation (5 attempts failed and needed to be repaired at
the Bruker factory).

Krishna
TIFR-Hyderabad.

On Tue, Jun 27, 2023 at 3:43 AM Angelis, Anna De via groups.io <adeangel - ucsd.edu_at_groups.io> wrote:

> Hi Tianzhi.
>
> In 2018 we moved 5 magnets at UCSD (500 to 900 Bruker, Varian and Magnex)
> using Bruker. They subcontracted to San Diego Machinery Movers for the
> rigging and Doug Lyons from Bruker energized and cryoshimmed them all back
> up without any issue. Doug is the best.
>
> Anna
>
>
> On Jun 26, 2023, at 6:21 PM, Shin, Ronald <shinr_at_uab.edu> wrote:
>
>
>
> Another endorsement for ARK Machinery Movers. Bruker had contracted these
> guys to install and move the magnets, and under the supervision of one of
> their magnet engineers (Doug Lyons), and they did a fantastic job. They're
> professional folks, courteous, and more than willing to answer the
> multitude of questions I had for them.
>
>
>
> We moved our Bruker Ultrashield (Magnex built) 500 magnet, and we also had
> a Bruker Ascend 600 and 850 installed as part of this package.
>
>
>
>
>
>
>
> Ronald Shin
>
> University of Alabama at Birmingham
>
> Central Alabama High Field NMR Facility
>
>
>
>
>
>
>
> *From:* main_at_ammrl.groups.io <main_at_ammrl.groups.io> *On Behalf Of *Todd> Rappe via groups.io
> *Sent:* Monday, June 26, 2023 9:48 AM
> *To:* main_at_ammrl.groups.io
> *Subject:* Re: [AMMRL] NMR relocation
>
>
>
> All of the following info is from 2011, so I don't know if it is still
> relevant or not:
>
>
>
> John Davidson is a former Oxford engineer and is an excellent resource. He
> installed our Oxford 800 in the early 2000s and successfully de-energized
> it and moved it to Florida in 2011.
>
>
>
>
> ARK Machinery Movers is the company Bruker used to deliver a 700, 850 and
> 900 to our lab and move a 700 and we were very happy with them. They have
> the equipment, personnel and expertise to safely move this large equipment.
> Our contact was Rich Kilduff , 443-386-6017, ark2103_at_aol.com
>
>
>
>
> _____________________________________
> Todd Rappe
> Minnesota NMR Center <nmr.umn.edu
> https://nmr.umn.edu/
> 612-625-9147 (office)
> 651-321-TODD (rings office and mobile)
> 612-624-8892 (magnet lab)
>
>
>
>
>
>

--

--
Krishnarao Doddapuneni, PhD
Tata Institute of Fundamental Research-Hyderabad
Survey No.36/P, Gopanapally Village
Serilingampally Mandal, Ranga Reddy Dist.
Hyderabad 500046
India
Office:040 20203053


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#243): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/243__;!!PvDODwlR4mBZyAb0!VyDKu_rgoV7MzzA2S0YIrc0qwQ1iVcPq=
-Z4kdX-3uL2n9Tl3VZ5yER7diG0QjeUy614w4apEjmXOqqowGgb7jNz3$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99788707/=
7559972__;!!PvDODwlR4mBZyAb0!VyDKu_rgoV7MzzA2S0YIrc0qwQ1iVcPq-Z4kdX-3uL2n9T=
l3VZ5yER7diG0QjeUy614w4apEjmXOqqowGvQFLj03$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jun 27 2023 - 04:13:11 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST