From: Bob Berno <bberno_at_mcmaster.ca>
Date: Thu, 22 Jun 2023 15:54:14 +0000
Hi Ken,

You may need to update the FIELD value in the edlock table. I know this was the case for older hardware/software, and may still be the case for your AV3.

Open edlock, and you should see the value for field. This is the value that the spectrometer expects for CDCl3.
Next, load a sample in CDCl3 such as the ethylbenzene standard and lock.
Once locked, check the value for field in the BSMS display. If it is significantly different from the value in the edlock table, then update the FIELD value in the edlock table (there should be a button or menu to update the FIELD). You may be prompted to enter the NMR administrator password.

Then load a sample in methanol -d4 and try to lock. It should now lock on the CD3 signal instead of the OD.  If not, then you may need to change another entry in the edlock table, but try this first.

Good luck,
Bob. 


Dr. Bob Berno
Manager: McMaster NMR Facility
Department of Chemistry & Chemical Biology
McMaster University, ABB-156
1280 Main Street West
Hamilton, ON L8S 4L9
(905) 525-9140 ext:24158
Fax: (905) 522-2509

From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Sent: Thursday, June 22, 2023 11:33:06 AM
To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
Subject: [AMMRL] MeOD lock issues on Avance III TS3.6.5 #Bruker
 

Ok, recently all of my MeOD samples are failing to lock on my Prodigy TCI in an Avance III running TS3.6.5.  These samples lock fine on other instruments.

I can manually set the field and it will lock, but as soon as it locks on another solvent with another sample, it's unable to apparently find the field automatically with MeOD. As far as I can tell, the lock phase is ok. 2H tune is spot on.

Any suggestions?
_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Online (#2= 33) | Reply To Group | Reply To Sender | Mute This Topic | New Topic
Mute #bruker
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Thu Jun 22 2023 - 08:54:18 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST