From: Hansen, Alex <hansen.434_at_osu.edu>
Date: Thu, 22 Jun 2023 15:36:09 +0000

Is your field value near its limit (-9999)? You may need to change your base
frequency. Although I typically only see this if a sample is locked onto
the wrong solvent.


[The Ohio State University]
Dr Alexandar L Hansen
Lead Research Scientist
CCIC & Gateway NMR Facility
R381 CBEC, 151 W Woodruff Ave, Columbus, OH 43210
614.292.7133 Office / 716.341.8476 Mobile
hansen.434_at_osu.edu<mailto:hansen.434_at_osu.edu>

Pronouns: he/him/his

"Who's ready to make some science?" - Cave Johnson

From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Kenneth Sharp-Knott
Sent: Thursday, June 22, 2023 11:33 AM
To: main_at_ammrl.groups.io
Subject: [AMMRL] MeOD lock issues on Avance III TS3.6.5 #Bruker

Ok, recently all of my MeOD samples are failing to lock on my Prodigy TCI in
an Avance III running TS3.6.5. These samples lock fine on other instruments.
I can manually set the field and it will lock, but as soon as it locks on
another solvent

Ok, recently all of my MeOD samples are failing to lock on my Prodigy TCI in
an Avance III running TS3.6.5. These samples lock fine on other instruments.

I can manually set the field and it will lock, but as soon as it locks on
another solvent with another sample, it's unable to apparently find the field
automatically with MeOD. As far as I can tell, the lock phase is ok. 2H tune
is spot on.

Any suggestions?



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#232): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/232__;!!PvDODwlR4mBZyAb0!SpX8L9vX35EFO8LPUanzDnKbGl7fpH8N=
qYDs4RuaOs2bvInHs9wSJyA6o1VCzNX98ZQnpUWuuIUvgMGF5XY$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99700062/=
7559972__;!!PvDODwlR4mBZyAb0!SpX8L9vX35EFO8LPUanzDnKbGl7fpH8NqYDs4RuaOs2bvI=
nHs9wSJyA6o1VCzNX98ZQnpUWuuIUvecWV--I$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!SpX8L9vX35EFO8LPUanzDnKbGl7fpH8NqYDs4R=
uaOs2bvInHs9wSJyA6o1VCzNX98ZQnpUWuuIUvJRuUejY$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jun 22 2023 - 08:39:30 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:42:24 MST