Dear Nadia
Topspin 1.3 pl10 is the last version of software that can be used with a DRX
console. The most "modern", but also out of support, OS is CentOS5.
Newer operating systems have not been validated with this vintage of hardware.
Your instrument was produced in 1997 and we ended the sales of DRX consoles
in 2002 and provided official support until 2016.
Sorry
Clemens
On 6/8/2023 10:29 PM, nadia beepath wrote:
HI everyone,
Good day!
I'm currently running TOPSPIN 1.3 on a Windows XP operating system with a
400 MHz Avance DRX Console.
1. I know that when this was installed in 2008, TOPSIN1.3 was the highest
version which could be used with the DRX Console. Has this changed and is
it possible to use a newer version of TOPSPIN with the DRX Console?
2. If the answer is no to Question 1, can TOPSPIN 1.3 (Windows or LINUX) be
installed with a newer operating system?
I will appreciate hearing from anyone who has more experience with this or
who has tried to upgrade their software with older Avance systems.
Thanks in advance. Have a great day!
Best regards,
Nadia
Nadia Singh (PhD)
Research Fellow
Chemistry Department
The University of the West Indies
St. Augustine, Trinidad and Tobago
Tel: (868) 662-2002 ext. 84053 / 85246/ 83542; (868) 795-4844
Email: nadia.singh_at_sta.uwi.edu<mailto:nadia.singh_at_sta.uwi.edu> / nadiabeepath_at_hotmail.com<mailto:nadiabeepath_at_hotmail.com>
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#186): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/186__;!!PvDODwlR4mBZyAb0!VTnacYHFXxaPGbyhSd24K2DyjxdHgSiw=
J8Hv25hz76nsNXFrFcP6pqZmiIR_oTb1_n-XpFlY1OV53CE08RXrrFTz8fjv$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99421420/=
7559972__;!!PvDODwlR4mBZyAb0!VTnacYHFXxaPGbyhSd24K2DyjxdHgSiwJ8Hv25hz76nsNX=
FrFcP6pqZmiIR_oTb1_n-XpFlY1OV53CE08RXrrCnL3CtC$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Jun 09 2023 - 05:03:55 MST