Re: [AMMRL] SampleCase loading errors

From: Bob Berno <bberno_at_mcmaster.ca>
Date: Thu, 27 Jul 2023 10:20:58 -0400

Hello Jessica,

You may wish to nudge up the supply gas pressure in the console to help the manual eject.
We had to do that with one of our magnets equipped with a Samplecase
that failed to eject when operating in manual mode.  I had tried to
increase the gas flow in the BSMS panel as had been suggested, but that didn't work.
But the pressure regulator inside the console was reading 5 bar.  I nudged
it ever so slightly to just above 5 bar (maybe 5.2 bar, see photo at the following
link:
https://photos.app.goo.gl/A76q7ewJihw2RSSf8
and the sample ejected fine.

Good luck,
Bob.

On 2023-07-26 10:11 p.m., Roberts, Jessica via groups.io wrote:
>
>
>
> Hi everyone,
>
> Thank you all for the detailed responses - I appreciate the guidance
> as I work through this!
>
> A brief update - a student was utilizing the instrument tonight and,
> while the sample went down into the magnet (after a few up and down
> cycles), it appears TopSpin does not realize there is a sample within
> the magnet. Troubleshooting thusfar has involved disabling the
> autosampler and trying to utilize the lift option to manually eject.
> Unfortunately, it appears the air pressure is not ample enough to
> eject the sample/spinner.
>
> Does anyone have any tricks for this issue? I will be working on this
> an hoping to resolve this as soon as possible. (and then get back to
> addressing the other potential troubleshoots for the previous email's
> issues).
>
> Thanks again everyone!
> Jessica
>
> Dr. Jessica M. Roberts, Ph.D.
> Assistant Professor of Instruction
> NMR Facility Manager
> Department of Chemistry and Biochemistry
> Ohio University
> 245 Chemistry
> Athens, OH 45701
> 740.593.0038
> jroberts_at_ohio.edu
> ------------------------------------------------------------------------

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
----------------------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon" wampum agreement.



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#375): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/375__;!!PvDODwlR4mBZyAb0!U5E_BJEfeA-KDQO4DU3_yJGl4R7Qcuv5=
rELvD9amHyXsgiiS-DHVOuAnR8mmnoffI-bQhO5AKgL4iey3$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100377317=
/7559972__;!!PvDODwlR4mBZyAb0!U5E_BJEfeA-KDQO4DU3_yJGl4R7Qcuv5rELvD9amHyXsg=
iiS-DHVOuAnR8mmnoffI-bQhO5AKhdJZg_C$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jul 27 2023 - 07:21:06 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:02 MST