Re: [AMMRL] SampleCase loading errors

From: Casey Simons <csimons_at_uoregon.edu>
Date: Wed, 26 Jul 2023 20:50:07 +0000

Hey Jessica,

As far as problem number 1. We had a similar situtation. It ended up being
an dirty shim stack. I started with cleaning the spin jets on the top of the
shim stack and this did not solve the problem. I ended up taking the head off
the shim stack and sonicating it in ethanol/methanol and a lot of junk was
dislodged. My hypothesis was uneven airflow pushing the sample slightly off skew.

Problem 2. Sometimes the autosampler does not see the position sensor. At
a previous position I made a little flag for the sensor to recognize a little
easier.

C

________________________________
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Roberts, Jessica <jroberts_at_ohio.edu>
Sent: Wednesday, July 26, 2023 11:54 AM
To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
Subject: [AMMRL] SampleCase loading errors

Hello everyone!

I have currently been having issues while using the SampleCase autosampler
on my 500 MHz Ascend instrument. There are two major issues that arise:

1. SampleCase fails to load a sample into the magnet. Autosampler will retrieve
the sample, send it up and across and down into the magnet. After a few seconds,
the sample will be ejected (often times by a sound that seems alot like building
pressure in an air line). The sample will then sit with the lift on for a few
seconds before repeating the process. After a few failed attempts the sample will
error out on IconNMR.

2. There is a less frequent issue but I will include it in case it is relevant.
I have seen the SampleCase retrieve a sample from the carousel, lift it to the
top track where it remains momentarily, then it will be sent back down to the
carousel. This typically times out after a few repeatedly failing attempts.

A few notes I will say:

-I am located at Ohio University where it has been quite humid as of late.
My air dryer does not have an autoswitch but I have powered it on today and
intend to run it for a few days to see if it improves things.

-Initial issues were only observed with one ceramic spinner, so I thought it
might be a problem with the old reflective tape. Now it can occur with a spinner
that worked for the previous experiment. In the most recent instance, the loading
error did not occur for a standard proton but for the immediately following standard
proton with rotation.

-It sometimes seems sporadic. A user can submit samples without issues while another
will have problems and it is hard to reproduce as I try to troubleshoot.

-I have not yet had issues with manual sample loading using lift on/off on BSMS
module so I suspect it is autosampler related.

Thanks everyone!
Jessica


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#369): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/369__;!!PvDODwlR4mBZyAb0!Rub8yDTb96GYvjd2prIbmVBJ3tqw2CUd=
q6hThXi_9HwK2AlGyjuanTzCWDtlykdxVdxPLvmFTkshgBSsIw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100377317=
/7559972__;!!PvDODwlR4mBZyAb0!Rub8yDTb96GYvjd2prIbmVBJ3tqw2CUdq6hThXi_9HwK2=
AlGyjuanTzCWDtlykdxVdxPLvmFTktwXj9O3g$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 26 2023 - 13:50:25 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:02 MST