Re: [AMMRL] SampleCase loading errors

From: Banghao Chen via groups.io <chen=chem.fsu.edu_at_groups.io>
Date: Wed, 26 Jul 2023 20:42:46 +0000

Hi Jessica,

Assume that the probe/stack of your spectrometer are clean, and your user used
a quality NMR tube, fine spinner and proper amount of sample, those are the
issues resulted in a similar problem on our AVIII with samplecase. If you can
rule out above situatuions by cleaning and testing with different sample and
spinners. You can try to go to BSMS service via "ha", adjust the lift control
pressure and/or lift offset, it may resolve the problem. You can also do the
samplecase calibration. check the maximum time to wait for the spinner to settle
in the BST after the lift air has been turned off. the minimum is 3s, default 20s,
maximum 60s for our system.

Good luck!

Banghao Chen/FSU

________________________________
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Martha Morton <mmorton4_at_unl.edu>
Sent: Wednesday, July 26, 2023 3:39 PM
To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
Subject: Re: [AMMRL] SampleCase loading errors

It might be an alignment issue. We had an alignment problem with our Sample
Case when we moved from Solid-State to liquids. There is a dummy load in the
sample case tools. You might talk to Center about how to do this process.

________________________________
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Michael Groves <m.groves_at_utah.edu>
Sent: Wednesday, July 26, 2023 2:19:55 PM
To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
Subject: Re: [AMMRL] SampleCase loading errors

Hi Jessica,

I have a few thoughts about your samplecase issues.

For issue 1 - The three attempts to insert the sample is standard behavior
and usually happens when the spinner gets stuck at the interface between the
shuttle that transports the sample and the top of the shim system. If you're
lucky, it might just be one (or a few) spinner that's having problems. In that
case it's an easy fix and you just take that spinner out of service. If it's not
the spinner, then frequently the sample changer has drifted out of alignment with
the top of the shim system. In that case you have to re-align it, but that's not
trivial and if you don't get it right, it can degrade the performance of the sample
changer further. I would start by pulling problematic spinners out of service
and see if that helps things.



For issue 2 - I've seen this one time and it was with a user who was using Jae-Young
tubes with dark colored valves on top. The sample would rise up into the sample
transport and the valve of the tube would be detected by the sample changer instead
of the spinner. So the system closed the latch that holds the sample up before the
spinner was in position. When it turned the air off, the sample dropped back into
the sample holder. I'd check to see if something is tripping the sensor that detects
the spinner in the top of the sample transporter prematurely.

Hope this makes sense.

Cheers,

Mike


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#368): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/368__;!!PvDODwlR4mBZyAb0!VKzY7ASmGOMzGv59Q1iCAUi7nwzw9Cs6=
KkLkZmXgd6Y5YqcpFZFY21qXKVUweiFCQkUjhdvitldiqmktN8Whg3Z8DU9z$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100377317=
/7559972__;!!PvDODwlR4mBZyAb0!VKzY7ASmGOMzGv59Q1iCAUi7nwzw9Cs6KkLkZmXgd6Y5Y=
qcpFZFY21qXKVUweiFCQkUjhdvitldiqmktN8Whg4zb6HE9$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jul 26 2023 - 13:43:20 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:02 MST