Re: [AMMRL] Overcurrent SCB20 Shim Board #Bruker #Hardware #Troubleshooting

From: Ulrich <Ulrich.Haunz_at_uni-konstanz.de>
Date: Thu, 6 Jul 2023 14:28:25 +0200

Hello Adil,


I had also too much troubles with my old AVANCE DRX 600 - the BSMS got
me crazy - overheating error.

The problem was easily solved - there was the fan try under the BSMS
unit and there was a filter in between. And this filter was closed with
dust. I cleaned this one and it was fixed. This dust filter was under
the BSMS unit and you have to put off the power chord to get access to
this plate.

At the NEO the fans are on the top of the rack.

Please look into the webinterface of the BSMS unit. There is something
about this special board.

There are some limits set for the ampere values also for the watts
running through the shim system.

This will heat up the whole device - so the current is limited there.


Please remove the shim system and controle all the connections and blow
the dust out of the whole system.

I think you have trouble in the shim system and this ramps the current
up there.


Yours sincerely,


Ulrich Haunz


Am 06.07.23 um 13:04 schrieb Muhammad Adil Raees via groups.io:
> Dear All,
>
> One of the SCB20 shim current board in AVANCE NEO console of our 600
> MHz NMR instrument shows error messages as below:
> /Shim current source 13 shutdown due to over temperature or broken
> connection! Reboot SCB20 (2)
> /1014: Coil power limit reached: PV = 23W
> /Shim SCB20 Handle Interrupt: SCB20 Current Source Overcurrent Error!
> All current zeroed: Board:1
>
> In the beginning, rebooting the machine helped in recovering the board
> but now, it hardly recovers and remains in the error state. Bruker's
> service engineer remotely checked the system and suggested replacement
> for this board. The fans in the BSMS rack are working well and the
> door filter is clean too. I am curious if the fault can really be with
> the board or if there is something else which is affecting its
> performance. Has anyone ever experienced the same or similar problem?
> Are these advanced electronic boards repairable?
>
> Thank you.
> Adil Raees
> --
> Muhammad Adil Raees, Ph.D.
> Manager NMR Facility
> The Central Laboratory
> Syed Babar Ali School of Sciences and Engineering
> Lahore University of Management Sciences
> Opposite Sector-U, D.H.A. 54792, Lahore
> +92-330-2866755
> adil.raees_at_lums.edu.pk
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#282): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/282__;!!PvDODwlR4mBZyAb0!WMMsHV3YARANm_BdUp3uYKmBp-LobDOI=
yvd9dI2it3EjYDmgviJgoF0fY3iSeEgqZB2qEE-KsTXia-8oxZp7QPAvZ_55aA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99984259/=
7559972__;!!PvDODwlR4mBZyAb0!WMMsHV3YARANm_BdUp3uYKmBp-LobDOIyvd9dI2it3EjYD=
mgviJgoF0fY3iSeEgqZB2qEE-KsTXia-8oxZp7QPBxGM18ZA$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!WMMsHV3YARANm_BdUp3uYKmBp-LobDOIyvd9dI=
2it3EjYDmgviJgoF0fY3iSeEgqZB2qEE-KsTXia-8oxZp7QPC56ZA4KA$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!WMMsHV3YARANm_BdUp3uYKmBp-LobDOIyv=
d9dI2it3EjYDmgviJgoF0fY3iSeEgqZB2qEE-KsTXia-8oxZp7QPDG39AlSw$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!WMMsHV3YARANm_BdUp3u=
YKmBp-LobDOIyvd9dI2it3EjYDmgviJgoF0fY3iSeEgqZB2qEE-KsTXia-8oxZp7QPCQvNRlmQ$=
 
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jul 06 2023 - 05:28:35 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:02 MST