Dear Hsin,
Please do what you suggested. Clone the probe, give it a new name, and clone
a MAS spinning profile and use that new probe name for identification. This
way all works just fine.
Contact me should you have further questions.
Cheers,
Jochem
Jochem
https://www.bruker.com/en/landingpages/bbio/world-class-bruker-nmr-trainings-2023.html
Jochem Struppe, Ph.D. Senior Application Scientist 2 Solid State NMR
Bruker BioSpin Corp.
15 Fortune Drive
Billerica, MA 01821
Phone: +1 978-667-9580 Ext.5113
Mobile: +1 -978-973-3547
Jochem.Struppe_at_Bruker.com<mailto:Jochem.Struppe_at_Bruker.com>
https://www.bruker.com
________________________________
-Bruker Confidential-
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Hsin Wang
Sent: Wednesday, July 5, 2023 3:03 PM
To: main_at_ammrl.groups.io
Subject: [AMMRL] Cloning a probe in TopSpin
I need to clone a probe in TopSpin. Specifically, I have solids probes tha=
t can be tuned in double resonance mode or triple resonance mode and the pu=
lses/powers are slightly different. I thought if I can clone a probe to a =
new probename, I may edit a new prosol table. but the they should be assoic=
ated with some other common characteristics, such as mas spinning profile.
Can this be done?
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#280): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/280__;!!PvDODwlR4mBZyAb0!W4PbNGMj9h1YNEYUiQfVOvl8MiyeDHWU=
KH9WR3qq7J-jHMdi3MJBjqQ0vwVorbBFV7o_PNPxKbUt2U3rlsjfERYoJOth$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99974068/=
7559972__;!!PvDODwlR4mBZyAb0!W4PbNGMj9h1YNEYUiQfVOvl8MiyeDHWUKH9WR3qq7J-jHM=
di3MJBjqQ0vwVorbBFV7o_PNPxKbUt2U3rlsjfEX2qlNoh$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Jul 06 2023 - 05:18:11 MST