AMMRL: Drying gas for low temp experiments

From: Stephen Philson <philson_at_nmr.chem.umn.edu>
Date: Fri, 30 Jul 2010 12:21:36 -0500

We have run into problems with the house nitrogen we normally use for
low (and high) temp work. It is provided throughout the two buildings
the chemistry department occupies, and is generated by boiloff from
a large storage tank of liquid nitrogen beside one of the buildings.
Some time ago we began to find rather large amounts water in the nitrogen
line, resulting in blockage of the external heat exchanger coils
(sitting in LN2) on one instrument, and ice in the probe on another
(where the low temp gas was from LN2 boiloff from a dewar in the lab,
but spinning and eject was still the house N2).

A recent change of LN2 suppliers seems to have reduced--but not
elimated--the problem. Since we don't really understand what is
happening, we are considering the installation of some sort of dryer in
the N2 line for the lab. Since the house N2 is almost certainly drier
than the house compressed air, and since we use it much less often
(the low or high temp work is a very small part of our operation), it
seemed to me that drying N2 within the lab might be a reasonably simple
solution to our problem. Of course, the modest and infrequent use might
make some of the normal drying solutions (like automatically regenerated
dessicants) problematical.

The questions I have are:

1) Has anyone else seen this kind of problem, or does anyone have a
plausible explanation for it? The storage dewar was recently replaced,
and as part of the process the dew points at our outlets where checked;
they seemed to be within specs, whatever that meant. It seems to me they
were something like -17C, but I don't really remember, and I think anyway
we need lower than they were likely to be able to measure. I presume we
have a leak somewhere, and with pipes and outlets all over two buildings
tracking it down is very difficult. We started having trouble before
the storage tank replacement, not much for a while, and now it has been
showing up again.

2) What sort of dryer would be best, and are there any recommended
sources? I've done a little investigating, but I'm having a little
trouble getting hold of the right person to talk to. It's been awhile
since I've been to an ENC where some of the appropriate vendors might
have been represented.

3) How much do we need to worry about oxygen at high temp? We do use
N2 for high temp operation, to avoid probe degradation, but if the N2
has water it may also have O2. How much is too much?

4) Perhaps getting our own dewar for LN2 boiloff would be a better option.
Does anyone have a idea of how long a dewar of LN2, set up for gas usage
and not liquid, would last if only used infrequently? And how long when
used continuously for one or two instruments?

Sorry for such a lengthy post. I welcome any suggestions.

Steve
-- 