Re: AMMRL: Varian autotests

From: <jerryh56_at_purdue.edu>
Date: Mon, 13 Jul 2009 11:35:29 -0400

Hi Tara,
I don't know what hardware you have, but here are a couple of shots in the
dark, considering the basic conditions in your console. Flaky results are very
often due to unfavorable operating environment, rather than hardware failure.
First, I would ensure every fan in the console is running well. Then pull the
shim power supply board and clean it (take antistatic precautions). The Inova
consoles I know are notorious for dust buildup on the shim power supply. Also
clean the filters in the front doors of the console. That should make a good
physical environment for the electronics.
To ensure a good electronic environment, check the power supply voltages with a
multimeter, and check each channel's DC balance from the amplifiers, and check
any gradient amplifiers' DC balances. DC output from all amps should be 0.00V
with power on, but no output. If there is any significant voltage of either
polarity, adjust it to 0.
I can't give you any advice regarding Varian's Autotest. We don't use it. I
find it easier to troubleshoot the console directly, rather than interpret
their test results.
Best wishes,
Jerry Hirschinger; NMR Instrumentation Specialist
Purdue University Interdepartmental NMR Facility


Quoting Tara Sprules <tara.sprules_at_mcgill.ca>:

> Hi,
>
> I've been running the varian autotest utility the last several months as
> was suggested to help find the source of some lock (?) problems I've been
> having. What I found however, is that each time I run autotests, a
> different subset of them fail (images, cancel, phasestab90ch1,
> pstab90Gch1, lockpower and lockgain). Images is the only one that
> consistently fails, the others vary. The same thing happens with both a
> cold probe and an rt probe. I suspect there may be something flaky
> somewhere in the hardware, which would explain why the autotest results
> are inconsistent, but I also wonder how reliable they are.
>
> The one thing I haven't done is run them repeatedly on my other
> instrument, which is functioning perfectly well. The one time I did
> lockgain and lockpower failed (so I wonder how useful those are as I don't
> think there is anything wrong with it).
>
> Does anyone have any thoughts on the utility of autotest? Or what may be
> causing the specific failures that I see? It seems images would suggest an
> imbalance in the quadature, although the only things I have been sent to
> replace so far were the lock XCVR, lock controller and lock preamp.
>
> Thanks,
>
> Tara
>
> The spectral symptom that I get is ringing in 2D homonuclear experiments
> that is sometimes really bad, and sometimes barely there. Lab environment
> is stable, legs are floating properly when I check...
>
>
> *********************************************
> Dr. Tara Sprules
> Quebec/Eastern Canada High Field NMR Facility
>
> www.nmrlab.mcgill.ca
>
> phone: (514) 398-1721
> fax: (514) 398-8254
>
> 3420 University St., Rm 023
> McGill University
> Montreal, QC, H3A 2A7
> *********************************************
>
>
Received on Mon Jul 13 2009 - 05:37:59 MST

This archive was generated by hypermail 2.4.0 : Fri Jun 16 2023 - 15:40:22 MST