Hi,
I've been running the varian autotest utility the last several months as
was suggested to help find the source of some lock (?) problems I've been
having. What I found however, is that each time I run autotests, a
different subset of them fail (images, cancel, phasestab90ch1,
pstab90Gch1, lockpower and lockgain). Images is the only one that
consistently fails, the others vary. The same thing happens with both a
cold probe and an rt probe. I suspect there may be something flaky
somewhere in the hardware, which would explain why the autotest results
are inconsistent, but I also wonder how reliable they are.
The one thing I haven't done is run them repeatedly on my other
instrument, which is functioning perfectly well. The one time I did
lockgain and lockpower failed (so I wonder how useful those are as I don't
think there is anything wrong with it).
Does anyone have any thoughts on the utility of autotest? Or what may be
causing the specific failures that I see? It seems images would suggest an
imbalance in the quadature, although the only things I have been sent to
replace so far were the lock XCVR, lock controller and lock preamp.
Thanks,
Tara
The spectral symptom that I get is ringing in 2D homonuclear experiments
that is sometimes really bad, and sometimes barely there. Lab environment
is stable, legs are floating properly when I check...
*********************************************