RE: AMMRL: Air-cooled compressor replacement.

From: Hirschinger, Jerry D <jerryh56_at_purdue.edu>
Date: Fri, 27 Jan 2017 22:52:25 +0000

Hi Matt,
The issue with a rebuilt compressor is whether or not the entire compressor capsule was replaced, or only the oil was replaced. Bruker has them rebuilt by replacing the entire capsule, so there is no possibility of contamination in the oil. A standard rebuild only replaces the oil, so contamination may remain.

We tried a rebuilt compressor from Cool Pair Plus - a company which by all accounts does excellent work. The cost was around $3000 and I installed it myself as you suggest. It only lasted a year before it caused problems. It was replaced by a Bruker rebuild which has not been a problem ($15K plus installation). You can get one brand new from Bruker for about $5K more.

IF I could find a distributor to sell a NEW Sumitomo, then I might try it.

Cheers, -Hirsch
Jerry Hirschinger, NMR Instrumentation Specialist
Purdue Interdepartmental NMR Facility
560 Oval Dr.  West Lafayette, IN   47907-2084
Office:           Wetherill 365A
Phone / Fax:      (765) 494-5288 / 494-0239
Cellular:         (765) 427-3034



-----Original Message-----
From: Matthew Cliff [mailto:matthew.cliff_at_manchester.ac.uk]
Sent: Friday, January 27, 2017 5:08 AM
To: ammrl_at_ammrl.org
Subject: AMMRL: Air-cooled compressor replacement.

Hi,

I read with interest the email summary from Bob Berno about cryoprobe compressors. In this summary it is implied that it is possible to source the air-cooled compressor independently of Bruker and install it yourself. This would be a Sumitmo FA-70 outdoor unit to replace a CNA-61 unit, for a Bruker cryoprobe. If anyone has experience of doing this, please could you let me know how it went? I suspect all that is needed is to plumb and wire-up the unit as in the old one, and then do some "warm cycles".

Best wishes,

Matt


--
----
Dr. Matthew J. Cliff
Senior Experimental Officer
NMR Facility
Manchester Institute of Biotechnology
Princess St.
Manchester
M1 7DN
Tel 0161 3065179.


Received on Fri Jan 27 2017 - 12:53:46 MST

This archive was generated by hypermail 2.4.0 : Tue Jun 20 2023 - 13:24:06 MST