Re: AF-80 electromagnet

Robert Taylor (taylor@mail.chem.tamu.edu)
Wed, 10 Feb 1999 11:59:23 -0800

Hi!

Let's be optimistic and assume that there is no hardware problem.
On Aspect 3000 systems with a SCM unit for the shims, there is
sometimes a problem after installing a new version of software. For
the shims, the software creates a SHIMATRX.999 file which may cause
problems with executing shim commands in the new version of software.
As an easy attempt, you might try deleting the SHIMATRX.999 file and
letting the new version of software create its own.

Good luck.

Bob

Dan F. Nogales wrote:

> Hi All,
> We have an AF-80 instrument that had some problems so we
> decided to make a few changes, mainly because we received
> some parts from a neighbor school Boise State.
> 1. We put in a new digitizer boar into the Aspect 3000. This
> solved the major problems of not being able to scale the gain etc.
> 2. We installed the new upgraded hard drive from Computer Field
> Services. We did this because the old HD crashed in the process
> of figuring out the digitizer problem.
> 3. The new HD will not accept the older ADAKOS software so I
> loaded the new stuff ADAKOS.
> 4. I also loaded the new DISB94 software
>
> The problem that we are having is the shim computer crashes
> when any buttons are activated. I think this is a communication
> problem with the software. Are these new versions of the software
> compatible with my older instrument?
>
> The new hard drive will not allow me to load an older version of
> ADAKOS, so I tried the new Version of ADAKOS with DISB86 and
> the same problem occurred. So is this a problem with ADAKOS
> or am I barking up the wrong tree?
>
> Any help is greatly appreciated. Dan Nogales (208) 467-8672
>
> xxxxxxxxxxxxxxxxxxxxxxxxxx
> Daniel F. Nogales Ph.D.
> Department of Chemistry
> Northwest Nazarene College
> Nampa, ID 83686
> (208) 467-8672
> dfnogales@science.nnc.edu
> http://science.nnc.edu/~dfnogales/
> xxxxxxxxxxxxxxxxxxxxxxxxxxxxx

--
Robert E. Taylor, Ph.D.            E-mail:  taylor@mail.chem.tamu.edu
Department of Chemistry            Voice:   (409) 845-8993
Texas A&M University               Fax:     (409) 845-4719
P.O. Box 30012
College Station, TX  77842-3012

URL: http://wwwchem.tamu.edu/services/NMR/