AMMRL: New 120 Hz spikes in proton spectra

From: Charles F. Wandler <wandler_at_chem.wwu.edu>
Date: Thu, 25 Feb 2010 13:49:31 -0800

Folks,

A 120 Hz spike has appeared in our Varian Inova 500,
and is especially large in samples like 1%H2O/D2O,
where to the left and right of the H2O blips appear
every 120 Hz, some in phase, some out-of-phase.

Any ideas on what this might be? We have had construction
under our 500 instrument, but I see nothing out of the ordinary.

Upon looking on that Internet thing, one site mentioned:
Because power lines carry 60-Hz ac, the voltage on them passes
through two peaks each cycle (one positive and one negative) and
passes through zero twice each cycle. This gives 120 peaks and 120
zero crossings in each second. Power-line noise follows this pattern,
generally occurring in bursts at a rate of 120 (sometimes 60) bursts
per second. This gives power-line noise a characteristic sound that
is often described as a harsh and raspy hum or buzz. Because the
peaks can occur twice per cycle, true power-line noise usually has a
strong 120-Hz modulation.

Or perhaps it could be something going out in the Inova console?

Any ideas or thoughts are most welcome.