Dear Spinners,
Following up the questions I asked a few weeks ago regarding what
UPSs are suitable to Bruker Avance consoles, especially for the
solid-state ones, I summarize here what I have found. Depending on
the configuration of consoles, different kinds of UPSs have to be
used:
1. Solution NMR consoles: (per console)
~5 KVA, single-phase/208V, cost: ~$5k
2. Solid-state NMR consoles with one 1KW high-power amplifier: (per console)
~10 KVA, 3-phase/208V, cost: ~20k
3. Solid-state NMR consoles with 2 1kW high-power amplifiers: (per console)
~15 KVA, 3-phase/208V, cost: ~22k
Please contact me for more details. By the way, I would like to thank
the people who replied to my inquiry.