Re: air dryers

From: Michael Strain <strain_at_mango.uoregon.edu>
Date: Wed, 1 Mar 2000 10:31:42 -0800 (PST)

On Wed, 1 Mar 2000, Joseph wrote:

> Colleagues
> We are considering getting a Balston model 75-20 (-100F dewpoint)
> compressed air dryer. Does any one out there have any comments (good or
> bad) they are willing to share regarding these dryers? Do they need a
> earlier stage (e.g. -40F dewpoint) to work properly?
> Thanks for your input.
>

We have two of the Balston 75-20 units... they work fine but
a ballast tank on the outlet is essential to even out pressure
spikes when the unit cycles between the two columns. We use
the Balston 72-012 "Receiver Tank".

If your house air is particularly icky an extra trap is a good idea.
Our house air is pretty oily.

Each Balston 75-20 will support more than one NMR system.
We have our 300 and 500 connected to one unit.

--Mike
-----------------------------------------------------------------------