Hi Sheng,

 

If you are using VnmrJ, after you set "tn=Co59", you will also need to set
tof=561980 [or sfrq=118.5] for a 500 MHz system to see the signal of 0.5M
K3CO(CN)6 in D2O.

This corresponds to tof=450000 [or sfrq=94.8486] for a 400 or tof=680000 [or
sfrq=142.223] for a 600.

 

The natural linewidth is ca. 6 Hz; any additional linebroadening you see
will be due to temperature gradients. The line moves about 170 Hz per
degree C on a 500 MHz system.

 

This sample is so sensitive that you can see the signal even if the probe
isn't tuned exactly. (I've seen the signal even when the probe was detuned
by 6 MHz.)

 

Contact me if you have more questions about this 59Co sample.

-Paul

 

Paul A. Keifer, Ph.D.

 <mailto:pakeifer_at_gmail.com> pakeifer_at_gmail.com

650-533-0212

 
<http://cp.mcafee.com/d/1jWVIq40Ui43qbbNEVud7bX5QjtPoVBNx5YQsTdzDTbCzBx5ASep
sohvd7a9J6XbybabWb1Lo8hYNOJAaCfff1i7P-QOZoQVsSxY_JcLmdendLvWbO2r_nV5dZVWUXHT
bELccsyqenXT76zDBHFShjlKCVOEuvkzaT0QSyrhdTV5MQsL8EIFTudTdw0zVga-xa7bVa6bqSul
-jLNhdHql-jLN1h4jVsSyCqekQPob6Azh0dWh1FEwS21EwEmFfRjh1hrXitBrpOH2xKVKsIyrh-B
qB5F> www.linkedin.com/in/paulkeifer1

_at_paulkeifer1

 

 

 

From: Cai, Sheng [mailto:sheng.cai_at_marquette.edu]
Sent: Thursday, December 11, 2014 1:14 PM
To: ammrl_at_ammrl.org
Subject: Re: AMMRL: Co59 NMR

 

Hi,

 

Thank you all for your suggestion. The sample I am using is Co59 standard
K3CO(CN)6, which should give a sharp peak (line width 5 Hz) at 0 ppm. So
large line width or huge chemical shift range should not be problem here.
And I can see good tuning signal.

I called Agilent and they suggested I used the wrong 1/4 ware cable on the
low band. My 94 MHz Co59 signal cannot pass through this cable. This
surprised me since I can run Si29 (79 MHz), C13 (100 Mhz) and P31 (160 MHz)
without any problem.

I ordered a new 1/4 wave cable anyway. I will let you know the results after
I tried it again with the new cable.

 

Sheng

 

 

  _____

From: Cai, Sheng
Sent: Wednesday, December 10, 2014 4:06 PM
To: ammrl_at_ammrl.org
Subject: Co59 NMR

 

Hi,

 

Did anybody run 59Co NMR before? I tried everything and it didn't work. I
even got no signal from the standard K3Co(CN)6, which should give a sharp
peak. My tuning is good and I am sure my probe is OK. Any trick I might be
missing? Thanks.

 

Sheng Cai

Marquette University

Milwaukee, WI 53233