RE: AMMRL: Issue with Lock window

From: Derrick C Kaseman <dckaseman_at_ucdavis.edu>
Date: Tue, 11 Apr 2023 23:52:53 +0000

Hi all,
Thank you for the many suggestions. At the end of the day the display cable
on the BSMS had wiggled loosed and was fixed by unplugging and re-plugging
the cable back in after 10s with topspin closed. When topspin was re-opened
the lock signal appeared normal.

Thanks!
Derrick


Dr. Derrick C. Kaseman
Technical Director of the NMR Facility
dckaseman_at_ucdavis.edu<mailto:dckaseman_at_ucdavis.edu>
530-752-7794<tel:+15307527794>
Use our Facility<https://www.nmr.ucdavis.edu/frequently-asked-questions#Training>

From: AMMRL List <ammrlrev_at_webserver2.chem.hawaii.edu>
Sent: Friday, April 7, 2023 9:27 AM
To: ammrl_at_ammrl.org
Subject: AMMRL: Issue with Lock window

Hi all,
I have a bruker Avance II system that I need assistance with. I ran a cf after a board swap (2H-TX). CF completed without error, but the lock window will not initialize in topspin (v2.1). When I put in the lock command the BSMS light turns on to indicate that the sample is locked and I can perform shimming. Therefore it appears that the lock is working, I just can't see anything in the lock window from topspin.
Any ideas?
Thanks,
Derrick