Dear All,

I would like to thank a lot all the many members who answered me so quickly. By the way, I fully apologize for the delay of my answer: I had some tricky delays avoiding me to check the advises on my spectrometer.

So, in summary, here is a procedure to change the administration password in Topspin (running on Windows) :

        - login as “nmrsu"
        - on the desktop, double click on “Bruker Utilities”
        - click on “Command Prompt”
        - run the following commands on the same line:
                        perl\bin\perl.exe prog\bin\installnmr nmrsu
        - here, you will be invited to enter the former password and then the new one

And it works!!

Thanks again to all the members who answered me!