Re: VNMR 6.1C installation on Inova 500

From: Sandip Sur <sur_at_chem.rochester.edu>
Date: Wed, 11 Apr 2001 11:26:22 -0400

Dear members,

The problem was in the 'config' file as correctly pointed out by Sara Kunz
and David Jones. Thanks to them for replying to me on the same day. I also
received reply from KRISHNAMURTHY_KRISH, Rolf Kyburz and Dr Toshi Nishida.
I checked the config file immediately and caught the problem right away and
I was ready to go. I was initially overconfident about the config file
which I read from the console (./setacq) during the installation and
checked the config file again before I started using the spectrometer. I
also ran 1d n15-h1 hsqc that uses the third channel. It worked okay
convincing me that the config file was set up correctly and there may be
other things that needed to set up. It may be that while I was checking the
config file after the initial set-up, I accidently changed the rf
configuration on the 2nd channel.

For such problem in this type of rf configuration:

1. check the 'config' file
2. check 'rfchannel' for its existence in the current experiment. You may
have rfchannel set up incorrectly. If you don't need 'rfchannel', destroy
it in the experiment. If it is a global parameter, destroy it.
3. check rfchannel switching by running h1 nmr with tn=h1 dn=c13 and tn=h1
and dn=h1 combinations
4. If all fails, blame it to the relay.


Thank you all very much,


Sandip



>Sandip,
>The problem is probably with your configuration file. You probably need
>to have U+
>Direct Synthesis on teh Ch 1 and U+ H1 only on Ch 2. This is how our 2
>channel
>Inova 400 is setup. By any chance to you have a printed copy of the
>config from the
>earilier version of the software?
>Good luck.
>
>Sara
>
>Sandip Sur wrote:
>
>> Hi Everyone,
>>
>> We replaced the old Ultra 1 (running 5.2F) with an Ultra 10 /6.1C /Solaris
>> 8 on our INOVA 500.
>>
>> We had to do recabling on the MSR board for the lock/shim to work.
>>
>> We also have all band 1st CH and H1 only 2nd CH. We noticed that these two
>> channels are not switching anymore on its own as they used to do with VNMR
>> 5.2F. The rfchannel command works fine. Therefore, the relay that
>> switches the channels is working okay.
>>
>> Do anyone know what will make the channel switch automatically depending on
>> the nucleus on Channel 1 and two or we just have to use rfchannel command
>> for each experiment requiring proton thru Channel 2 and BB thru channel 1?
>>
>> Thanks,
>>
>> Sandip
>>
>> *** BEST WISHES FOR 2001 ***
>>
>>
>>________________________________________________________________________________
>> Sandip K. Sur, Ph.D. Email:
>>sur_at_chem.chem.rochester.edu (W)
>> Manager and Application Scientist
>> NMR Facility
>> Department of Chemistry Tel.: (716) 275 4705 (W)
>> B53B Hutchison Hall (716) 275 0912 (W)
>> River Campus (716) 586 7794 (H)
>> University of Rochester Pager: (716) 275-4231 (ask for me)
>> Rochester, NY 14627-0216 Fax: (716) 473 6889
>> Monday, April 9, 2001
>>
>>________________________________________________________________________________
>
>--
>
>Sara Kunz
>Chemistry Dept. MS 015
>Brandeis University
>PO Box 549110
>Waltham, MA 02454-9110
>Phone: 781-736-2840 Fax: 781-736-2516



                        *** BEST WISHES FOR 2001 ***

________________________________________________________________________________
Sandip K. Sur, Ph.D. Email: sur_at_chem.chem.rochester.edu (W)
Manager and Application Scientist
NMR Facility
Department of Chemistry Tel.: (716) 275 4705 (W)
B53B Hutchison Hall (716) 275 0912 (W)
River Campus (716) 586 7794 (H)
University of Rochester Pager: (716) 275-4231 (ask for me)
Rochester, NY 14627-0216 Fax: (716) 473 6889
Wednesday, April 11, 2001
________________________________________________________________________________
Received on Wed Apr 11 2001 - 14:34:17 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 04 2023 - 15:18:03 MST