another summary of Oxford AS magnet drifting

From: Wei Li 448-7532 <wli_at_utmem.edu>
Date: Fri, 06 Apr 2001 15:25:32 -0500

Hi friends,

Thanks to so many people who take their time responding to my question:
high lock z0 drift of our Oxford AS-500 magnet with Inova console. Dr.
Chung from Scripts provide an archive from an earlier discussion for a
600MHz, the address is

http://wwitch.unl.edu/ammrl/archives/Nov_Dec-99/0100.html

http://wwitch.unl.edu/ammrl/archives/Nov_Dec-99/0106.html

To summarize the responses I got:
1. For many magnets (the majority responses) from Varian, one has to
reset the lock frequency due to the drifting. The resetting time ranges
from 3 months to a year depending on the drift rate. One usually does
not need to do that in Bruker magnet. The reason might due to different
design in the console rather than the magnet.

2. Most responses indicate that there will be no big problem in terms of
daily operation, although it is an inconvenience. However, probe tuning
and changed values from other shims(i.e. lineshape) may need to be
monitored due to the change of z0 value.

3. Most responses indicate that frequency drop out of filter
bandwidth/amplifier is unlikely to happen before reseal the magnet in
10+ years. A good suggestion is try to see if one can get a free
servicing pump up from Varian/Oxford.

4. Since my quoted magnet drifting rate is <10Hz/hr, there is no solid
base to question Varian. In fact, it is not too bad to have 5~6 Hz/hr
drifting.

Based on the responses, I think I can relax and try to work something
out with Varian/Oxford.

Cheers,

Wei Li
-----------------------------------------