Re: [AMMRL] Pulsar benchtop

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Mon, 5 Oct 2026 19:17:12 +0200
Hi Rob,

can you possibly borrow an oscilloscope somewhere to check the respective outputs of the frequency
generator/synthesizer, pulse generator/transmitter and - carefully, at first with a -30 dB
attenuator - the amplifier? Usual outputs of synthesizer and transmitter of (*insert NMR manufacturer
company name here*) NMR consoles are usuallyin the 0.5…3 V range. If you find only a few mV here,
that’s the likely issue.
RF amplifiers normally need said attenuation to not fry the scope. From the found peak-to-peak
voltage you can (web-) calculate the output in Watts, which should be somewhere in the 20-100 W
range at full power…

Cheers
Bert

-> Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers



_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Onl= ine (#3822) | Reply to Group | Reply to Sende= r | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
--Apple-Mail-77EFAE9D-70BF-4A74-8829-E367FDA2EC87-- Received on Mon Oct 05 2026 - 10:37:31 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 16 2026 - 09:11:46 MST