Re: Sn199 NMR

Juergen Schulte (schulte@binghamton.edu)
Fri, 14 Feb 1997 08:24:16 -0500 (EST)

On Thu, 13 Feb 1997, Qing Ning wrote:

> Hi! there,
> I was trying very hard to run a Sn119 (II) spectrum, but nothing observed
> so far. This is the Sn(Ac)2 family. I used Sn(CH3)4 as the standard, a very
> nice singlet peak was obtained in 4 scans. However, when I put my sample
> in, and sweeped from -2000 ppm to +2000 ppm, I can not find any peak. Is
> there anyone who knows Sn(II) NMR better? Any kind of commentwill be
> helpful.
>
> Qing Ning
>
> ----------
> Qing Ning, Ph.D
>
> NMR Specialist Analytical Service Lab
> q-ning@nwu.edu Department of Chemistry
> tel. 847-491-7080 Northwestern University
> Fax 847-491-7713 2145 Sheridan Road
> Evanston, IL 60208
>
>
>
>

Hi,

I believe, that Sn-119 has a negative magnetogyric ratio. If you use
proton decoupling you might lose the signal due to the negative NOE.
Try Inverse Gated Decoupling or maybe a DEPT instead.

Juergen

---------------------------------------------------------------------