Re: Bruker AC monitor - summary
Paul J. Kanyha (info@tecmag.com)
Thu, 5 Sep 1996 15:33:35 -0500
>Thanks to all those who responded to my question about Bruker AC monitor
>(Aspect 3000). I have enclosed a summary below. I am trying the second
>solution.
>
>
>solution 1: contact MR Resources (800-443-5486) (same monitor costs ~$1400),
>solution 2: replace the monitor with simple CGA color monitor such as an
>NEC Multisync II, model JC-1402HMA (used one, ~$200), after building an
>adaptor. Wiring of the connector adapter:
>
> 8-pin connector (Bruker) DB-9 connector (NEC)
> 1 open -
> 2 3
> 3 4
> 4 5
> 5 open -
> 6 1
> 7 8
> 8 open -
>
>
>=================================================================
>Quincy Teng, Ph.D. Office: (706)542-1502
>Director, NMR Facility Lab: -9314
>Department of Chemistry Fax: -9454
>University of Georgia teng@sunchem.chem.uga.edu
>Athens, GA 30602-2556
>=================================================================
Dear Dr. Teng:
If you are interested in purchasing a new CGA monitor, a Sony PVM-1390 13"
CGA monitor is available from B&H Photo-Video (800-947-9901) for $589.
Best Regards,
Paul J. Kanyha, Ph.D.
Tecmag, Inc.
6006 Bellaire Blvd.
Houston, TX 77081
T: (713) 667-8747
F: (713) 667-3180
info@tecmag.com
www.tecmag.com