The 400 at Frontier is an ECP400 not an ECX400. There are no swappable
components between the 2 models.

> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Casey Simons via groups.io
> Sent: Thursday, September 21, 2023 5:08 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] New NMR Director

Hey Steven,

Congrats on the position! If you are still rocking the ECX-300 and not an
ECZ-300 I would suggest grabbing the entire ECX-400 console/ATM unit/probes
(assuming this is the system from Frontier). There was also a posting earlier
from Chuck Smithheart at Delta State giving up ECX-300 parts. Might be worth
the freight bill to have all these on hand unless the faculty are planning
to write for a replacement.

C

________________________________
> From: main_at_ammrl.groups.io on behalf of Ashok Krishnaswami
> Sent: Wednesday, September 20, 2023 5:24 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] New NMR Director

Steven

The console that you are talking about is an even older model than your ECZ300
and there are no electronic components that are compatible.

There are two attachments that you could get from that system.

Autotune Controller as a backup to the one you have

8 position sample changer.

Regards

Ashok

> From: main_at_ammrl.groups.io On Behalf Of Steven Butler via groups.io
> Sent: Wednesday, September 20, 2023 3:54 PM
> To: main_at_ammrl.groups.io<mailto:main_at_ammrl.groups.io>
> Subject: [AMMRL] New NMR Director

Hello,

I am a new NMR director and need some advice. We have an older JEOL 300 MHz
Spectrometer in our department and a local industry company recently upgraded
their spectrometer from a JEOL 400 MHz system. They have offered to donate
any useful parts from the old system to us. Anyone familiar with JEOL able
to tell me what would be useful for me to pick up from them?

Thanks,

Steven Butler

Steven Butler
Instrumentation Specialist
Department of Chemistry and Biochemistry
Office: (435) 797-1606
WIDT 231



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#619): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/619__;!!PvDODwlR4mBZyAb0!Sc_HopgPQDEumgD01pjryHCWZL6xybIi=
LLSrSdpBS-9YGAxun2SOnx7vPYhzW2Wo9-qBDv5-zJU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101486488=
/7559972__;!!PvDODwlR4mBZyAb0!Sc_HopgPQDEumgD01pjryHCWZL6xybIiLLSrSdpBS-9YG=
Axun2SOnx7vPYhzW2Wo9-qB0YgiOic$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 22 2023 - 04:40:42 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:44:18 MST