Re: [AMMRL] Sharing one of our learning of Bruker Sample Express

From: Jennifer <jennifer08um_at_gmail.com>
Date: Thu, 14 Sep 2023 15:29:20 -0400
You will notice the rotor off position when you take off the cassette.

Jennifer 

On Sep = 14, 2023, at 2:25 PM, Cathy Clewett via groups.io <clewett=wisc.edu_at_gr= oups.io> wrote:



Can you also describe what symptoms you were having = before you found this problem?

 

Thanks!

 

Cathy

 

Catherine F.M. Clewett, PhD

(she, her, hers)

Senior Scientist, Magnetic Resonance Facility

Department of Chemistry, University of Wisconsin â= €“ Madison

1101 University Ave

Madison, WI 53706

clewett_at_wisc.edu

(608)262-8196

Pbcic.chem.wisc.edu

 

 

 

From: main_at_ammrl.groups.io <main_at_ammrl.gro= ups.io> On Behalf Of Joseph Sachleben
Sent: Thursday, September 14, 2023 12:57 PM
To: main_at_ammrl.groups.io
Subject: Re: [AMMRL] Sharing one of our learning of Bruker Sample Ex= press

 

Where exactly is this on the SamplExpress? Can you p= rovide more pictures?

 

Thanks

 

Joe S



On Sep 14, 2023, at 12:18 PM, Xiaoyan Sun via groups= .io <xsun9=ncsu.edu_at_grou= ps.io> wrote:

 

Dear AMMRL,

 

We would like to share one of our recent learnings=  about Bruker Sample Express with you all.  

Hope this piece of information will be helpful for y= our future needs.  

 

After five years of seamless operation, our Bruker S= ample Express has begun to exhibit subtle signs of aging. During a recent i= nspection, we identified a concerning issue: three of the motor placeholder= s within the autosampler displayed clear signs of wear and tear, and they were found to be completely broken = when we removed them as shown in attached photos.

 

After some research and invaluable assista= nce with Peter Demou, we found the correct name of the part for the space h= older of the motor is General Purpose Vibration-Damping Sandwich Mounts wit= h Stud and Insert. However Mcmaster does not have the right size as 8*8*6 3M.  https://www.mcmaster.com/products/b= umpers/metric-general-purpose-vibration-damping-sandwich-mounts-with-stud-a= nd-insert/

 

Don't be confused with the Load-Rated Threa= ded Bumper (htt= ps://www.mcmaster.com/93115K871/). It can not work because it does not have the top female portion.Â=  

 

Finally, We found the correct parts at amazon, let&#= 39;s see how long it will last.  

 

Have a wonderful day!

Jennifer

 

-- =

Jennifer Sun

Senior Research Scholar , Molecular Education, Techn= ology and Research Innovation Center

Joint Research Assistant Professor, Department of Chemist= ry

North Carolina State University

2620 Yarbrough Drive

Raleigh, NC 27607

(919) 515-1160

xsun9_at_ncsu.edu

https://research.ncsu.edu/metric/

 

Please citing METRIC to support us:

All NMRÂ=  measurements were made in the Molecular Education, Technology and Resear= ch Innovation Center (METRIC) at NC State University.

<IMG_2201.jpg><IMG_2196.jpg><= /p>

 

_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Online= (#587) | Reply To Group | Reply To Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Thu Sep 14 2023 - 17:43:48 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:44:18 MST