the shift of the resonance is proportional to the quality factor of the probes.
The higher the Q the worse it gets. So Prodigy probes will be affected but
not a much as helium probes.
The measured and corrected values refer tot he reading at the thermocouple
and a corrected value that is based on a correction that is entered by the
user. The correction is varying from probe to probe and is also dependent
on other factors such as airflow. The correction is a linear correction and
can only give a good fit over a limited range of maybe 30 - 40 degrees as
the large dependence of shift vs temperature is not linear
Clemens
On 9/7/2023 2:24 PM, Mark Swanson wrote:
**EXTERNAL EMAIL**
Hi Everone,
Does the issue with Ethylene Glycol also apply to N2 cryoprobes (e.g. Prodigy)
or is it mainly Helium cryoprobes? Also, under 'edte' there's a measured value
and a corrected value. Does anyone know how good the correction is in general,
or does it vary from probe to probe?
Thanks,
Mark
-=-=-=-=-=-=-=-=-=-=-=-