Re: [AMMRL] 99.8% MeOD vs 80% Ethylene Glycol

From: Clemens Anklin via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Thu, 7 Sep 2023 20:02:11 +0000

Hi Mark

the shift of the resonance is proportional to the quality factor of the probes.
The higher the Q the worse it gets. So Prodigy probes will be affected but
not a much as helium probes.

The measured and corrected values refer tot he reading at the thermocouple
and a corrected value that is based on a correction that is entered by the
user. The correction is varying from probe to probe and is also dependent
on other factors such as airflow. The correction is a linear correction and
can only give a good fit over a limited range of maybe 30 - 40 degrees as
the large dependence of shift vs temperature is not linear

Clemens


On 9/7/2023 2:24 PM, Mark Swanson wrote:
**EXTERNAL EMAIL**
Hi Everone,

Does the issue with Ethylene Glycol also apply to N2 cryoprobes (e.g. Prodigy)
or is it mainly Helium cryoprobes? Also, under 'edte' there's a measured value
and a corrected value. Does anyone know how good the correction is in general,
or does it vary from probe to probe?

Thanks,

Mark


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#552): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/552__;!!PvDODwlR4mBZyAb0!Xr3Dg3lwPby7TwbqD1sBWAJCqq07FJV7=
MwfUeld2qVxDhn4JBNWbLPb5O08f40YBeybYu_LP2uF8BUOvWS8WTlYpO2JF$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/100827188=
/7559972__;!!PvDODwlR4mBZyAb0!Xr3Dg3lwPby7TwbqD1sBWAJCqq07FJV7MwfUeld2qVxDh=
n4JBNWbLPb5O08f40YBeybYu_LP2uF8BUOvWS8WTn1c0ATE$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Sep 07 2023 - 13:03:01 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:44:18 MST