Dear Catherine,
In our new lab, we use Trigard and Toxgard II oxygen safety monitors from MSA (Mine Safety Apparatus). The Trigard is a single monitor for up to three sensors and we use it in our 9.4 T NMR lab, with one sensor at about 1 meter height for liquid nitrogen leaks, one sensor at about 3 meters height for liquid helium leaks and one sensor mounted directly on the Trigard monitor unit, on the wall next to the operator's desk. The Toxgard II is a single-channel monitor with a sensor mounted directly below the monitor unit on the wall. This sensor is attached via plastic tubing to a remote "sniffer" port at a height of about 2 meters inside the Faraday cage of our 7 T MRI scanner. Since the cryo-refrigerated MRI magnet contains only liquid helium, only one sensor is needed there. While the Toxgard II monitor contains an air sampling pump, the Trigard sensors rely upon passive diffusion to sample ambient air. More details are available at www.msanet.com<
http://www.msanet.com>.
Both the Trigard and Toxgard II sensors are electrochemical and require periodic recalibration and/or replacement. Our building maintenance staff check them once every 6 months. As long as these checks are done diligently, the monitors seems to be very reliable. They are wired into our building's automation system so that emergency ventilation fans turn on and I and others get a text if a low oxygen alarm goes off.
One final note: in our old lab, we used an Oxigraf O2iM laser oxygen monitor. This monitor is rather expensive but super-reliable. There are essentially no disposable parts inside and we only needed to have it repaired once in about 10 years of continuous service. Like the MSA Toxgard II, the O2iM contains an air sampling pump and can only monitor one location per monitor. For details, see http://oxygenanalyzers-oxygensensor.oxigraf.com/item/nema-4x-oxygen-analyzers/o2im-oxygen-deficiency-analyzer/07-0180?
Best regards,
Ken Fishbein
Kenneth W. Fishbein, Ph.D.
Facility Manager, MRI Section
National Institute on Aging
National Institutes of Health
251 Bayview Boulevard
Suite 100, Room 04B120
Baltimore, MD 21224
(410) 558-8512
FAX: (410) 558-8376
Kf31x_at_nih.gov<mailto:Kf31x_at_nih.gov>
From: Eugenio Alvarado [mailto:ealva_at_umich.edu]
Sent: Tuesday, September 23, 2014 2:43 PM
To: Clewett, Catherine; ammrl_at_ammrl.org
Subject: Re: AMMRL: Oxygen Monitors
We have used wall mounted oxygen sensors from GfG Instrumentation for about 5 years without problems or maintenance.
Eugenio Alvarado, Ph.D.
Co-Director, NMR Lab. Room 3500
Chemistry Department, University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924
On 09/22/2014 11:10 AM, Clewett, Catherine wrote:
Ammrl folks:
Does anyone have suggestions for rugged, inexpensive oxygen monitors? We will be installing an NMR magnet in a room with only one exit and no natural ventilation, so we thought it would be a good idea. What should I look for when purchasing one?
Thanks
Catherine F.M. Clewett, Ph.D.
Assistant Professor of Physics
Department of Mathematics, Chemistry and Physics
WTAMU Box 60787
Canyon, TX 79016-0001
Phone: 806-651-2545
Fax: 806-651-2544
Email: cclewett_at_wtamu.edu<mailto:cclewett_at_wtamu.edu>
Received on Wed Sep 24 2014 - 11:51:16 MST