Dear AMMRL (Probably mostly for Varian users),
I may be exposing ignorance that is not shared by others, but the Sun
Blade-100 on my Inova-500 started acting strangely (computer crashes for no
reason) over the past 2 weeks or so, and it finally died when I tried to
reboot it (IDPROM: Contents Invalid). The Ethernet MAC address and Host-ID
were are "F"s (FF:FF:FF...etc.).
I've since learned (thanks to Ken Kezeor at Varian) that these IDPROMs use
NVRAM, and actually have a battery embedded within the chip. When the
battery dies, the chip's memory is wiped, and your Sun is dead.
You can get these pre-programmed chips from Sun for $30 (or less if your
institution has a discount w/ Sun). The Sun part number (for both
Blade-100s and Blade-150s) is #100-6889. You can buy them online or through
a distributor. If you have these Suns, it might not be a bad idea to have a
spare on hand. Also, when you replace your IDPROM, the MAC address will
change as will the HostID. If you have software with license keys tied to
the HostID...well, you can see the problem. Supposedly, Sun will reprogram
a new IDPROM with your old HostID and Ethernet Address, but I don't know the
mechanism for having that done, or how much it would cost.
Yesterday I just swapped hard drives (and the second Ethernet card) with
another Blade-100, and I was back in business shortly. This would be a
bigger problem if you don't have spare Blades, and if you don't have a spare
IDPROM chip.
My Sun Blade-100 was just under 3 years old, and I think these are supposed
to last longer than that. This is just a word of warning (or, one more
thing to worry about).
Also, the Blade-1500s use EEPROMs instead of NVRAM chips for their IDPROM;
therefore, this shouldn't be an issue.
Cheers,
-Rich Shoemaker
---