if you change a Bruker SGU, you have to ensure, that the firmware of
all SGUs within your machine is equivalent. Using UniTool this
requires a long time but nevertheless this job is easy.
Bot *don't forget* to select the option "Init SGU" after the
firmware upgrade. This initialization is a *must* and cannot be
replaced by a reset of the AQS rack or even by switching off
and on the AQS rack.
If you don't choose the initialization you get very strange
mistakes of any kind.
I lost two days of my life .....