Re: Water linewidth

From: S W Simon Sham <ssham44_at_uab.edu>
Date: Mon, 27 Sep 2004 11:29:45 -0500

Regarding the problem of H20 linewidth posted last week, most people got
a linewidth at half-height around 30Hz. One reported ~20Hz ,and another
a linewidth at half-height of 60-70Hz with Varian spectrometer.

Therefore, ~30 Hz is a typical linewidth that one gets at 600 MHz.

Hope this may be helpful.

Simon Sham
------------------------------------------------------------------------
-----------------