Re: [AMMRL] Bruker Topspin 4.2 vs 4.3

From: Abil Aliev <a.e.aliev_at_ucl.ac.uk>
Date: Mon, 30 Oct 2023 18:29:53 +0000

Would be good to know IF and HOW Bruker does alpha and beta testing of
their new versions of Topspin/Icon before releasing them.

I tested 4.2 on my laptop first and noticed something rather odd data
syncing within minutes. This, together with my previous experience with
Bruker software, was sufficient to postpone using 4.2.0 on
spectrometers. The plan was to wait for 4.2.2 or higher.

The most stable windows version to me was 4.0.6. In 4.0.7, Icon started
wiping *.set files with the list of samples queued and it was rather
embarrassing to email our users asking them to resubmit their samples again.

We have been using 4.1.3 for a while now and twice the *.set files have
become unreadable by Icon and I had to start a new queue and users had
to resubmit their long kinetics experiments. The confidence level in
switching to even 4.1.4 is very low.

Abil


On 30/10/2023 16:04, Kenneth Sharp-Knott via groups.io wrote:
>
> ⚠ Caution: External sender
>
>
> I have seen a lot of comments from users who have said they 'rolled
> back' to 4.2 after numerous issues with 4.3.
>
> I would appreciate it if you could elaborate on what those issues are
> specifically.  Bruker would like the feedback as they are releasing a
> new version soon and want to make sure those issues are taken care of.
>
> One individual stated that they believed there to be more issues with
> 4.2 than 4.3 so they are surprised by this feedback.
>
> Best
>
> --
> Ken Sharp-Knott
> Manager of Analytical Services and the NMR Facility
> Department of Chemistry
> Virginia Tech
>
> (540)267-6502 (Cell)
> (540)231-0885 (Office)
>
--
Department of Chemistry, University College London
20 Gordon Street, London WC1H 0AJ
Telephone: 020 7679 4616 (internal 24616) and 07474 000 924
https://www.ucl.ac.uk/chemistry/people/abil-aliev
https://www.ucl.ac.uk/nmr


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#744): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/744__;!!PvDODwlR4mBZyAb0!UrbpsYWXoq4jmQax8VnoSXRAOg9xfB4c=
tTgZ9MQJ4RWA5UQxs_CRw0gBIjIdbVVeR_9ZmWFCbwCKNnpivinQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102278454=
/7559972__;!!PvDODwlR4mBZyAb0!UrbpsYWXoq4jmQax8VnoSXRAOg9xfB4ctTgZ9MQJ4RWA5=
UQxs_CRw0gBIjIdbVVeR_9ZmWFCbwCKNvpkq7tI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 30 2023 - 11:31:36 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST