Re: [AMMRL] Bruker AQS3+

From: Michael Groves <m.groves_at_utah.edu>
Date: Mon, 23 Oct 2023 15:29:45 +0000

For what it’s worth, it’s a tachometer signal and if the speed
drops too low you’ll get an alert. This is what causes problems with
some fans—if you get a fan with a slightly higher current draw the
fan won’t spin as fast and you’ll get alarms.

Cheers,
Mike

> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Peter Thompson via groups.io
> Sent: Saturday, October 21, 2023 8:34 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Bruker AQS3+

I believe the third line on those fans is a high-voltage/locked rotor alarm.

Peter

On Sat, Oct 21, 2023 at 3:16 PM Ulrich Ulrich.Haunz_at_uni-konstanz.de wrote:

Dear all!

Am Freitag, Oktober 20, 2023 20:34 CEST, schrieb "Sham, Shing" <ssham_at_luc.edu<mailto:ssham_at_luc.edu>>:

> Igor, thanks for pointing out the fan issue.
> In the past, when the fans literally died, the AQS server of my other magnet
> would tell me. I "wish" they could show the spinning speeds on those running fans.
> I bought some from an Amazon vendor, yeah I know, for the other magnet.
> They were cheaper but did not work. It ended up I had to buy from Bruker.
> Does anyone know where I can get them and they work?
>

> I also tried to save money - so I asked the distributor of these fans. This
> company made these ones especially for Bruker. So I had to order these from there.

> There are three lines - the third one is a special one - but I don't know for
> what reason this is - tacho, status or something else.

> The best buy is here at Bruker.

> Yours sincerely,

> Ulrich Haunz


> Thanks
>
> Simon
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#720): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/720__;!!PvDODwlR4mBZyAb0!RFd1JQiTdbgGW0kX01bJWREAirTUFb1G=
lStTOuodwF9TVbg5w81PrJSjFke2lhhHxhzAFgWzV6N17uwJ1g$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102082269=
/7559972__;!!PvDODwlR4mBZyAb0!RFd1JQiTdbgGW0kX01bJWREAirTUFb1GlStTOuodwF9TV=
bg5w81PrJSjFke2lhhHxhzAFgWzV6NUHFjEZQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 23 2023 - 08:29:52 MST

This archive was generated by hypermail 2.4.0 : Wed Nov 01 2023 - 14:23:34 MST