Unless I am missing something here, you absolutely CAN transfer the integrals
limits from one dataset to another…
In the dataset filename/exptno/pdata/1 (for example), after you have integrated,
there is a file “int2drng”. Just copy that into any other
dataset’s pdata directory and it will be used when you type “int”
(or “int2d”).
Hope this helps — again, forgive me if it was either obvious or not
what you meant!!
Best!
John R.
On Oct 18, 2023, at 10:58 AM, Hsin Wang <hwang1_at_ccny.cuny.edu> wrote:
Hi, George,
I asked Bruker’s hotline the same question recently. This is what
they told me and it works for me.
After you go through picking the integration, save it, exit integration, and
click the “INTEGRALS”, you will see the table you picked,
and the absolute values of your integrals.
It would be nice if one can transfer the integration limits from one experiment
to the next. But I have not figured that out. For now, I just typed in the
limits to get the consistency.
Hsin
> From: main_at_ammrl.groups.io On Behalf Of Sukenick, George
> Sent: Tuesday, October 17, 2023 7:02 PM
> To: main_at_ammrl.groups.io<mailto:main_at_ammrl.groups.io>
> Subject: [EXTERNAL] [AMMRL] integration between spectra
I have a researcher who wants to compare integration between different spectra
(all taken with same parameters).
My thought was to integrate without normalizing or referencing the integral,
but it seems that integration (at least in Topspin) does not provide the
absolute value.
I don’t see anything in the AU program collection either.
Is there a way to obtain this information?
Or while we always use relative integrals – is there a reason not to
use the absolute values?
I suppose that the data could be imported to a spreadsheet and then absolute
intensities summed within a region, but there should be a better way?
Thanks!
George Sukenick, Ph.D.
Associate Lab Member/Head
NMR Analytical Core Facility
Sloan Kettering Institute
Memorial Sloan Kettering Cancer Center
g-sukenick_at_ski.mskcc.org<mailto:g-sukenick_at_ski.mskcc.org>
=========================
Please note that this e-mail and any files transmitted from
Memorial Sloan Kettering Cancer Center may be privileged, confidential,
and protected from disclosure under applicable law. If the reader of
this message is not the intended recipient, or an employee or agent
responsible for delivering this message to the intended recipient,
you are hereby notified that any reading, dissemination, distribution,
copying, or other use of this communication or any of its attachments
is strictly prohibited. If you have received this communication in
error, please notify the sender immediately by replying to this message
and deleting this message, any attachments, and all copies and backups
from your computer.
Disclaimer ID:MSKCC
---------------------------------------------------------------------------=
-----------------------------------------------------
John Ralph
Department of Biochemistry, and the D.O.E. Great Lakes Bioenergy Research Center
The Wisconsin Energy Institute, Room 2129
University of Wisconsin, Madison
1552 University Avenue
Madison, WI 53726-4084
(608) 890-2429
E-mail: jralph_at_wisc.edu
http://www.biochem.wisc.edu/faculty/ralph/
http://www.glbrc.org
https://energy.wisc.edu/about/energy-experts/john-ralph
ORCID: 0000-0002-6093-4521
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#703): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/703__;!!PvDODwlR4mBZyAb0!VWFeiRImGXPdyv74UAP72EbTiBvkqTMi=
MlnGmkuQlps433fm2HOCdsad0JKmBogQ51VGr-JkqIEKrLItEs1Vkjl84ho$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102029964=
/7559972__;!!PvDODwlR4mBZyAb0!VWFeiRImGXPdyv74UAP72EbTiBvkqTMiMlnGmkuQlps43=
3fm2HOCdsad0JKmBogQ51VGr-JkqIEKrLItEs1Vm7a7yH0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Oct 18 2023 - 09:26:56 MST