Hi again all,
The problem ultimately turned out to be at the cryoplatform end. I
connected the laptop to a newer cryoplatform that is normally controlled
via TopSpin. That communication worked fine.
So I warmed up the problem cryoplatform and reset it. I also took the
opportunity to update the firmware. Now it is working fine.
I think I caused the problem by connecting the laptop before realizing I
needed a driver for the USB/serial converter. The communication worked
briefly but then hung up.
Rather than rely on this 14 year old laptop for too much longer I'll put
a replacement together, now I know what is needed to make it work.
Thanks,
Phil.
--